OLYMPUSTJF-Q180V duodenoscope, SN: 270****.

Test result:

1. Image ok;

2. Light bundlestill ok;

3. Insert tubestill ok;

4. Biopsy tubewear;

5. Bendingsection, wires and helix tube wear;

6. Knob leakage.

Maintenance:

Suggest toreplace biopsy tube, bending section, wires, helix tube, knob and rubber. (Nowarranty for CCD)

Repair cost: $****